Biomedical Engineering Reference
In-Depth Information
Table 4.3 Representation of data from full-resolution CI measurements
Approach
Advantage
Application
API mass on each component
including non-sizing parts,
such as the induction port
Simplicity
OIP quality control
API mass on each size-
fractionation stage of the CI
Basic aerodynamic
size-related
information
OIP development and in
bioequivalence
API mass mathematically
linked to aerodynamic
diameter range of sizing
stages
Aerodynamic size-
related information
comparable from one
CI system to another
OIP development and in bioequiva-
lence; this approach is essential
for input to models predicting
respiratory tract deposition
Fig. 4.7 Representation of CI raw data by mass deposition in both sizing and non-sizing compo-
nents ( From [ 78 ]— used by permission )
Fig. 4.8 API mass on CI size-separating stages as an ordinal function of stage d 50 size: example
from ACI operated at 28.3 L/min ( From [ 78 ]— used by permission )
Search WWH ::