Figure 6.53  Proteins purification by size-exclusion chromatography.
is done. A secondary flow, miscible with the first one, flowing at a larger flow rate,
confines the first flow.
Let us analyze the case of flow focusing along a wall and denote Q 1 and Q 2
as the two flow rates, with Q as the total flow rate Q 1 + Q 2 and w 1 and w 2 as the
two channel widths occupied by fluid 1 and fluid 2, as schematized in Figure 6.59.
After a short establishment length, the new profile is a Poiseuille-Hagen quadratic
profile. Clearly, the relative width e = w 1 / w is related to the flow rate ratio Q 1 / Q .
Using (2.47) for the flow in a rectangular channel
Figure 6.54  Separation principle of substance traveling through a chromatography column.