Tabl e 8. 1
Summary of operating parameters of endomicroscopes for four (contact, confocal, multiphoton, and OCT) imaging modalities
Lateral
Working
resolution a
distance
Implemented in
Modality
.m/
NA
FOV (m)
(m)
Signal type
Imaging depth (m)
clinic
Contact
imaging
6-10.0
N/A
500-1,000
0
Fluorescence
Top surface
Yes
Confocal
0.5-3.0
0.3-1.0
100-500
200-500
Reflectance and
fluorescence
200 in vivo
applications
Ye s
Multiphoton
0.5-3.0
0.35-1.0
100-500
200-500
Multiphoton
fluorescence
Usually up to 500 in
the lab and 200
in tissue
Animal imaging/ex
vivo
OCT
5.0-15.0
0.05-0.20
Limited by scanning
range. Can be
10s of mm across
0.5-3 mm
Coherent reflectance
imaging
1-2 mm
Yes
a
While NA of optics, confocal, and multiphoton endoscopes allow high lateral resolution, the sampling might be limited by density of the fiber bundle or
sampling of the scanner. The most important requirement is light collection for these two modalities
 