Biomedical Engineering Reference
In-Depth Information
Table 7.14: Example FMEA
Rating
Remedial
1
Display shows insulin
is being infused but no
insulin is entering the
body.
Patient can
suffer diabetes
related episode.
A flat battery
causes pump
to fail while
maintaining
display activity.
4
4
5
80
Occurrence
Detect = 5
1
2
3
4
5
1
5
10
15
20
25
2
10
20
30
40
50
3
15
30
45
60
75
4
20
40
60
80
100
5
25
50
75
100
125
Before
Occurrence
Detect = 1
1
2
3
4
5
1
1
2
3
4
5
2
2
4
6
8
10
3
3
6
9
12
15
4
4
8
12
16
20
5
5
10
15
20
25
After
Rating
Remedial
1
Display
shows insulin
is being
infused but
no insulin is
entering the
body.
Patient
can
suffer
diabetes
related
episode.
A flat battery
causes pump
to fail while
maintaining
display
activity.
4
4
5
80
Low battery warning
(buzzer) included to
activate when 2 hours
battery life remaining;
emergency battery
pack supplied; charger
designed to be portable
(handbag / pocket size).
4
2
1
8
 
Search WWH ::