Biomedical Engineering Reference
In-Depth Information
0 or 1 depending on whether
2 ( x , y ) is positive or negative. Therefore,
subtracting the two phase maps and adding 2
ϕ
1 ( x , y ) 2 ϕ
whenever the difference is negative
effectively yields a new phase map, which corresponds to the beat wavelength
π
Λ
12 .
Figure 7.11 shows the phase images of the USAF resolution target imaged at a slight angle.
The images produced with a single wavelength exhibit multiple phase steps (see
Figure 7.11A and B ). By comparing the phase images from each wavelength, the 2
π
phase
ambiguities can be resolved. For
λ
5 633 and
λ
5 532 nm,
Λ
5 3334 nm, which is high
1
2
12
enough to remove the discontinuities (see Figure 7.11C) .
The negative aspect of this method is that the phase noise is also amplified by the same
factor as the range. However, one can then use this dual-wavelength “coarse” map as a
guide, together with one of the original phase maps (
2 ), to produce the low noise
“fine” phase map [10] . If the noise in the coarse phase map is too excessive, some of the
single wavelength segments might still end up being vertically shifted from its correct
position by a single wavelength, creating phase image artifacts. These errors can then be
corrected in software by simply looking for such jumps and shifting them up or down [3] .
In comparison to the coarse map, the noise in the resulting fine map (see Figure 7.11D) is
much lower, while the axial range is still the same.
ϕ
1 or
ϕ
3.142
3.142
(A)
(B)
nm
nm
800
600
600
400
400
200
200
0
000
200
400
600
800
200
(D)
(C)
400
Figure 7.11
Phase maps for (A)
λ 1 5
532 nm and (B)
λ 2 5
633 nm; (C) 3D rendering of synthetic dual-phase
map with beat wavelength
Λ 12 5
3334 nm and (D) reduced noise fine map (the images are
174
μ
m
174
μ
m, 450
450 pixels).
3
3
Search WWH ::




Custom Search