Biomedical Engineering Reference
In-Depth Information
(C)
9
7
C b (t) (voxel)
5
3
C a (t)
1
-1
0
0.5
1
1.5
2
Time (min)
dC b (t)/dt = k 1 · CMRO 2 + CBF[k 2 ·C a (t) - k 3 ·n · C b (t)]
(B)
(A)
H 2 17 O Bolus Injection
17 O 2 inhalation
Fig. 15.4. Schematic diagram showing the multiple in vivo 17 O measurements for
determining CMRO 2 using the complete model according to the mass balance equa-
tion of Equation (15.8) which links Cb (t) ,Ca (t) , CBF and n with CMRO 2 . To simplify
the equation, three known constants of 2 α 1 , mf 2 and m/ λ used in Equation (15.8) are
replaced by k1, k2 and k3, respectively. ( A ) Stacked plot of the 17 O spectra of cere-
bral H 2 17 O tracer from one representative rat brain voxel as indicated by the circle in
the anatomical image (low center insert). The spectra were acquired before and after a
bolus injection of H 2 17 O for CBF measurements. ( B ) Stacked plot of the 17 O spectra
of the metabolic H 2 17 O from the same representative voxel acquired before (natural
abundance), during (as indicated by the gray bar under the stacked plot) and after a
two-minute 17 O 2 inhalation. ( C ) Measurement of Ca (t) by using an implanted 17 ORF
coil (the left insert). The middle insert illustrates a 17 O spectrum of natural abundance
H 2 17 O obtained from the rat carotid artery blood by using the implanted RF coil before
H 2 17 O inhalation. The right insert shows the time course of Ca (t) (circle symbol) and
Cb (t) from a representative 3D 17 O CSI voxel (square symbol) in the same rat during a
two-minute inhalation of 17 O 2 . Finally, the ratio between the 17 O signal decay detected
after a bolus injection of H 2 17 O (see Fig. 15.4A ) versus the 17 O signal decay detected
after the cessation of 17 O 2 inhalation (see Fig. 15.4B ) gives the constant of n. Adapted
from Zhu et al. of Ref (54) .
brain before the introduction of 17 O-labeled oxygen gas (see
Fig. 15.4B ) provides an excellent internal reference for quan-
tifying the absolute concentration of C b (t) for each CSI voxel
(54, 87) . This self-calibration method is independent of the 17 O
detection sensitivity. This is particularly valuable when a surface
RF coil with nonuniform spatial distribution of detection sensi-
tivity is applied.
The arterial input function of C a (t) wasmeasuredinvivobyan
implanted 17 ORFcoil (97) wrapped around a carotid artery (in
the rat). Figure 15.4C illustrates the implanted 17 O RF coil, the
3.2.3. Other
Measurements
 
Search WWH ::




Custom Search