Singh et al. [35] used suppressed ion chromatography to analyse low concentrations of
anions in non saline waters in the presence of high concentrations of matrix ions.
Table 12.9 gives a summary of various other methods described in the literature for the
determination of anions in non saline waters.
Fig. 12.23 Determination of anions in an aqueous sample
Conditions:
Eluant 1:
0.75mM NaOH
Eluant 2:
200mM MaOH
200mM H 3 BO 3
Eluant 3:
1.0mL min −1
Flow rate:
Gradient program
%1:
100 100 85 55 0 0
0
0 100
%2:
0
0 15 45 40 40 100 100
0
%3:
0
0
0
0
60
60
0
0
0
The next injection was always made 42min after the beginning of the
previous run
Source: Reproduced with permission from Dionex Corporation [24]
12.2.6 Mixtures of plant and soil extracts
12.2.6.1 Chloride, nitrate, sulphate and phosphate
Bradfield and Cooke [55] described an ion chromatographic method using an ultraviolet
detector for the determination of chloride, nitrate, sulphate and phosphate in water
extracts of plants and soils. Plant materials are heated for 30min at 70°C with water to