Biomedical Engineering Reference
In-Depth Information
Fig. 2.23 Trudell Medical International “delay” apparatus coupled to USP/Ph.Eur. induction port
with low-flow five-stage MMI developed for testing inhalation devices intended for use by infants
or small children
Vertical Moving Shutter Plate
VHC
Solenoid
MIC
OMRON
Timer
Reset
MIC
Power Solenoid
Control
Fig. 2.24 Schematic diagram of Trudell Medical International “delay” apparatus showing work-
ing components ( From [ 67 ] —used with permission )
elapsed time between MDI actuation and the onset of sampling is achieved by inter-
posing a mechanical shutter housed in a purpose-built adapter that locates on axis
between the VHC and induction port leading to a CI that is operated at a flow rate
appropriate for the intended patient category for the VHC (Fig. 2.24 ).