Biomedical Engineering Reference
In-Depth Information
Table 2.10
Values of
X
c
,
S
/
W
, Re
f
, and
d
50
(
μ
m) for the NGI size-fractionating stages at flow rates
(
Q
) associated with its archival calibration
All flow rates
Stage
1
2
3
4
5
6
7
MOC
a
X
c
N/A
b
N/A
b
0.35
0.41
0.61
0.84
0.85
0.94
S
/
W
1.0
2.0
3.0
3.0
3.0
3.1
4.9
7.1
Q
= 15 L/min
d
50
14.1
8.61
5.39
3.30
2.08
1.36
0.98
Do not use
Re
f
1,482
724
404
338
230
167
166
Q
= 30 L/min
d
50
11.8
6.40
3.97
2.30
1.36
0.83
0.54
0.36
Re
f
2,938
1,435
801
669
455
328
324
149
Q
= 60 L/min
d
50
8.06
4.46
2.82
1.66
0.94
0.55
0.34
0.14
Re
f
5,876
2,870
1,602
1,339
909
757
647
298
Q
= 100 L/min
d
50
6.12 3.42 2.18 1.31 0.72 0.40 0.24 0.07
Re
f
9,793 4,783 2,671 2,231 1,515 1,095 1,079 496
a
MOC
micro-orifice collector—used as an alternative to a backup filter; note that the size corre-
sponding to 80% collection efficiency (
d
80
) rather than
d
50
values are quoted for this equivalent to
a backup filter, rather than an impaction stage
b
Single-nozzle geometry; therefore, cross-flow parameter does not apply
In summary, there is some freedom in establishing the precise location of the
collection surface beneath the nozzle plate with many impactor designs except per-
haps the MSLI. This flexibility is available for the ACI and NGI, which are the
apparatuses most likely these days to be used in OIP aerosol characterization. More
importantly from a user's standpoint, small variations in the precise thickness
(depth) of a tacky coating that could be applied under normal circumstances to the
collection surfaces, to improve particle adhesion, are unlikely to affect stage perfor-
mance. As will be seen in Chap.
10
, the ability to mitigate bias due to particle
bounce and blowoff from the collection surface is particularly important with abbre-
viated systems when used to characterize both MDI- and DPI-generated aerosols.
The NGI represents the state of the art for CI design. Based on their experience
with the Marple-Miller series of CIs, the developers of the NGI took the opportunity
to ensure that the criterion for
S
/
W
was met for all stages (Table
2.10
). Likewise, the
multinozzle designs of stages 3-7 and the MOC were arranged such that values of
X
c
were significantly <1.2. Some values of Ref
f
are <500 for certain stages, particu-
larly those that size fractionate the finer particles and when operating this CI at the
lower flow rates within its range [
38
]. Nevertheless, all stage collection efficiency
curves for this impactor throughout the original design range of operation from 30
to 100 L/min were shown to be sharply defined with minimal overlap [
21
]. Even at
the lower flow rate extension at 15 L/min for nebulizer testing, these curves are still
acceptable for size-fractionating purposes [
22
]. Note, however, that the MOC
should be followed by a final filter or the internal filter option be used to replace the
MOC when used at 15 L/min [
22
].
Search WWH ::
Custom Search