Fig. 10.41
FPM
<5.0μm
for DPI product 1 determined by various FSI configurations in comparison
with NGI-measured APSD from measurements of Pantelides et al
.
(error bars represent ±1 SD for
three replicates) (
From
[
42
]—
used with permission
)
Fig. 10.42
Configuration of FSI for follow-on study by Pantelides et al. matching internal volume
more closely to that for the NGI (
From
[
45
]—
courtesy of D. Russell-Graham
)
The resulting pressure drop-elapsed time profiles for the same DPI product 1 as was
evaluated in the previous study are illustrated in Fig.
10.43
.
As expected from a comparison of internal dead volumes, the pressure drop-time
profile for the volume-enhanced FSI was much closer to that for the NGI. This
closer match of dead space resulted in a closer agreement in fine particle mass/actua-
tion between the two systems. Figure
10.44
illustrates the near-to-linear correlation
between fine particle mass and air acceleration rate for the FSI, volume-enhanced