Table 10.10
pMDI products evaluated by Guo et al
.
with the FPD-AVCI [
33
]
Formulation
type
Product name
API
Propellant
Excipients
Aerobid
®
Flunisolide
CFC
Sorbitan
trioleate
Suspension
Combivent
®
Salbutamol (AS)/ipratropium
bromide (IB)
CFC
Soya lecithin
Suspension
MaxAir™
Pirbuterol
CFC
Sorbitan
trioleate
Suspension
Advair
®
Fluticasone propionate (FP)/
salmeterol xinafoate (SX)
HFA
None
Suspension
Flovent-110
®
Fluticasone propionate
HFA
None
Suspension
Proair
®
Salbutamol
HFA
Ethanol
Suspension
Proventil™
Salbutamol
HFA
Oleic acid,
ethanol
Suspension
Atrovent
®
Ipratropium bromide
HFA
Water, citric
acid, ethanol
Solution
This pattern of divergence between the various abbreviated options was comparable
in absolute terms with the behavior observed with measures of
FPM
<4.7μm
. The fact that
the addition of Brij-coated filters to the collection dishes of the FPD-AVCI only
slightly reduced
EPM
<1.1μm
suggests that particle re-entrainment is an unlikely cause.
However, it should be emphasized that these differences between the abbreviated
systems with or without surfactant-saturated filters to control particle bounce and
re-entrainment and the control ACI were small, and therefore unlikely to prevent
any of these options being used as an abbreviated impactor of choice with this par-
ticular formulation. This type of detailed study illustrates well the approach that
should be taken when validating a potential AIM-based system for any OIP.
Guo et al
.
have also recently presented measurements undertaken with the
Westech FPD-AVCI impactor, evaluating eight different suspension and solution
pMDIs (Table
10.10
) [
33
].
CPF
>5.0μm
,
FPF
<5.0μm
, and
EPF
<1.0μm
determined by the
FPD-AVCI were compared to the same metrics obtained by means of an 8-stage
nonviable ACI. Ten actuations from the pMDI-on-test were delivered into the FPD-
AVCI or ACI, and API recovery proceeded afterward by validated quantitative
chemical analysis. Since the ACI does not have stages with cut-off diameters pre-
cisely at 5
m, the three measures of interest were interpolated from the
cumulative APSD data obtained with the full-resolution CI.
Equivalent total API recovery was observed for all pMDI products between FPD
and ACI (Fig.
10.26
). Although agreement between FPD-AVCI and ACI data was
generally good, Guo et al
.
found small but significant differences in all three
subfractions with some of the formulations, especially with
CPF
>5.0μm
, with the val-
ues obtained using the FPD-AVCI higher than their ACI-determined counterparts
(Fig.
10.27
).
Guo et al. concluded that, whether or not the pMDI is a solution or suspension,
formulation was not the only deciding factor on whether or not there is a divergence
between abbreviated and full-resolution data [
33
].
μ
m and 1
μ