Biomedical Engineering Reference
In-Depth Information
Fig. 10.13
Photomicrograph
of displaced Brij 35
surfactant on collection plate
for the second impaction
stage of the AIM-pHRT
impactor (
From
[
27
]—
used
with permission
)
Fig. 10.14
Comparative measures of impactor-sized subfractions by ACI with measures of CPF,
FPF, and EPF after modifying second impaction stage of AIM-pHRT CI with Brij-35-soaked filter
(
From
[
27
]—
used with permission
)
reduced systems. Despite this finding, measurement time gains were found to be
significant; analysis with the abbreviated systems taking around 30% of the time
consumed using the ACI. This finding of more rapid determinations is a consistent
trend across other reported studies to be discussed later in this chapter.