Fig. 10.5
Effect of number of actuations of Flovent-110
®
on C-FSA-measured data when used
with surfactant-coated collection plates (
From
[
19
]—
used with permission
)
Fig. 10.6
Use of liquid ethanol-sensitive paper to diagnose presence of liquid droplets within the
upper stages of the C-FSA. (
a
) Traces confirming liquid ethanol presence with incoming aerosol
to C-FSA from sampling 10-actuations of Qvar™-80, using liquid ethanol-sensitive filter paper
located on collection plate below stage. (
b
) Traces confirming liquid ethanol presence with incom-
ing aerosol to C-FSA from sampling 10-actuations of Qvar™-80, using liquid ethanol-sensitive
filter paper located on collection plate below stage 1 (
From
[
20
]—
used with permission
)
The introduction of additional dead space in the T-FSA, compared with the
C-FSA, was found to improve agreement with the ACI in terms of fine particle
mass, a result attributed to the provision of more similar conditions for ethanol
evaporation within the former configuration (Fig.
10.7
). Overall, the difference
between the data using the slightly modified C-FSA and T-FSA was so small that