Biomedical Engineering Reference
In-Depth Information
Table 7.1 Regression analysis and goodness-of-it statistics for LPM/SPM ratio versus MMAD
Optimum
boundary b (μm)
Average
MMAD (μm)
Coeficient of
determination, R 2 (%)
RMSE / b
(μm)
OIP ilecode a
Product type
CI runs ( n )
Slope ( b )
RMSE c
w9k201
HFA suspension MDI
80
4.7
3.91
0.4071
0.0162
96.4
0.040
w9j901
HFA suspension MDI
39
3.3
2.57
0.4959
0.0350
93.4
0.071
w9j801
HFA solution MDI
201
2.1
1.50
0.7155
0.0421
96.2
0.059
w9jk01
Dry powder inhaler
279
3.3
2.66
0.4319
0.0201
83.0
0.047
w9k901
Dry powder inhaler d
279
2.0
2.59
2.3831
0.1278
84.3
0.054
w9j601
CFC suspension MDI
43
2.1
2.54
2.4548
0.0872
95.5
0.036
w9k001
CFC suspension MDI
272
3.3
3.54
1.6127
0.0330
97.3
0.020
w9kw01
CFC suspension MDI
272
3.3
2.86
0.7046
0.0198
95.8
0.028
a The ilecodes are unique, randomly-generated alpha-numerical labels assigned to speciic products in the IPAC-RS database
b Here, optimum boundary is based on results from the eight-stage Andersen Impactor
c RMSE —root-mean-square error
d Modiied Andersen nonviable CI used for this product
 