Biomedical Engineering Reference
In-Depth Information
300
300
250
250
200
200
150
150
100
100
50
50
0
0
0
10
20
30
40
50
0
10
20
30
40
A
B
Time (s)
Time (s)
300
250
200
150
100
50
0
0
10
20
30
40
50
C
Time (s)
Figure 10.10
Binding of NH 3 in air to an optical fiber-based evanescent sensor. Effect of reversibility ( Cao and
Duan, 2005 ): (a) Run # 1 (b) Run # 2 (c) Run # 3.
Table 10.8: Binding rate coefficients for NH 3 in air to an
optical fiber-based evanescent ammonia sensor. Effect of
reversibility ( Cao and Duan, 2005 ).
Run #
k
D f
1
80.472
3.843
2.3350
0.0424
2
95.326
2.628
2.3902
0.02922
3
54.887 4.320
2.1132 0.07446
D f , respectively. The above numbers obtained from the fractal analysis do indicate that the
above analyzed runs are reversible. The error exhibited in the fractal dimension is relatively
small. Note that the fractal dimension is based on a log scale, and even small changes in the
fractal dimension indicate significant changes in the degree of heterogeneity that exists on the
biosensor surface.