Biomedical Engineering Reference
In-Depth Information
+9V
S1
R1
30.1K
BT1
9V BATTERY
IC2A
IC1
1
50Hz
5
3
3
2
CONTV
OUT
7
DISCH
CD4050B
2
6
8
RESET
TR IG
THOLD
RESET
IC3A
+9V
1
8
3
2
V+
D1
1N4148
ICM7555
1
CD4049
J1
C2
8
C1
0.47uF
_______
1
R2
1
f=
1.4RC
100, 2W
IC2B
IC3B
D2
1N4148
4uF
1
1
5
4
5
4
CD4050B
CD4049
8
8
C3
1nF
+9V
IC2C
IC3C
1
1
7
6
7
6
D3
1N4148
J2
C4
CD4050B
R3
8
CD4049
1
8
100, 2W
IC2D
D4
1N4148
4uF
1
IC3D
9
10
1
9
10
CD4050B
8
CD4049
8
IC2E
1
IC3E
11
12
1
11
12
CD4050B
8
CD4049
8
IC2F
1
IC3F
14
15
1
14
15
CD4050B
8
CD4049
8
Figure 8.48 Schematic diagram of a battery-powered fibrillator. A 555 timer IC generates a 50-Hz square wave. This signal drives paired
CMOS buffers. This output stage is actually a full bridge which causes doubling of the effective voltage across the heart.
draft form, they are very instructive as to the speci
fi
c methods applied to the design and
testing of implantable pacemakers and de
fi
brillators:
• prEN-45502-2-1, Active Implantable Medical Devices Part 2-1: Particular
Requirements for Active Implantable Medical Devices Intended to Treat
Bradyarrhythmia (Cardiac Pacemakers) , Final Draft, January 2001
• prEN-45502-2-2, Active Implantable Medical Devices Part 2-2: Particular
Requirements for Active Implantable Medical Devices Intended to Treat
Tachyarrhythmia (Includes Implantable De
fi
brillators), 1998
On the U.S. side, the FDA has not yet adopted a set of standards harmonized with
EN-45502. O
cially, the FDA still refers manufacturers to a guidance document written
in the 1970s:
• Donald F. Dahms, Implantable Pacemaker Testing Guidance, FDA FOD#383, posted
January 12, 1990, available at www.fda.gov/cdrh/ode/doc383.pdf
Search WWH ::