Figure 4.32 Setup for assessing the susceptibility of a medical device to magnetic fields. The magnetic field applied to three axes of the
device under test is 10 A/m. This test simulates interference that may be caused on a device by low-frequency magnetic fields, such as those
generated by the power lines.
1 meter
L1
PVC Elbow
J1
T1
Variac
T 2
1
5
PVC "T"
1 meter
To RMS
D VM
24V, 10A
AC Power Plug
110VAC
4
8
1
J2
#18 solid
insulated
wire
C
BN
2
R1
1 Ohm, 20W
3/4" PVC Pipe
Reference
Ground
Plane
Figure 4.33 A magnetic-field induction generator can be built from a coil of wire inside some PVC pipe. This assembly is driven by a
variac-transformer combination to produce the desired magnetic field at the center of the coil.
 