Biomedical Engineering Reference
In-Depth Information
L1
Hot
50uH
(Black)
C3
0.1uF 600VAC
C1
1.0uF 400VAC
R1
1K
Phase Selection Switch
J1
J2
To
Device
Under
Test
1
1
Hot
To
Powerline
Ground
3
3
Neutral
(Green)
2
2
R2
1K
Neutral
S SW DPDT
(White)
R3
50
C2
1.0uF 400VAC
C4
0.1uF 600VAC
L2
50uH
J3
BNC
1
2
To 50 ohm
Spectrum
Analyzer
Permanent Connection
to Ground
Figure 4.16 A 50 Ω/50 µH LISN as defined by standard CISPR16-1. This circuit provides a 50-Ω output impedance for measurement of
RF emissions produced by the device under test. Conducted emission measurements are carried out from 150 kHz to 30 MHz.
To Spectrum Analyzer
SA 1
SA 2
H-Field
Probe
5
8
L3
Metal Box
Insulated Wire
1
4
HOT
HOT
CB1
L1
500 uH
L4
50 uH
Black
Black
1
2
1
R1
C2
0.22uF 600 VAC
47
LA1
110V Neon Lamp Assy.
2
J1
R2
1
1
GROUND
GROUND
J2
Green
Green
3
3
C1
1uF 400VAC
C3
2uF 400 VAC
2
2
L2
500 uH
NEUTRAL
NEUTRAL
White
White
120 V
Figure 4.17 A safer way of running lab tests is to use a LISN in which radiated emissions are coupled inductively to a spectrum analyzer.
L3 is a single loop of No. 14 stranded insulated wire which is coupled to an H-
fi
eld probe made of a VCR head.
170
 