TABLE 4.2
EN-55011 Radiated Emissions Limits for Group 1 Devices a
Field Strength (dBµV/m)
Frequency (MHz)
Test Distance (m)
Class A
Class B
30-230
10
40
30
230-1000
10
47
37
a The lower limits apply at the transition frequency.
plane. The table is typically centered on a motorized turntable that allows 360° rotation. A
measurement antenna is positioned at a distance of 10 to 30 m as measured from the clos-
est point of the device under test. The radiated emissions are maximized by con
guring
and rotating the device under test as well as by raising and lowering the antenna from 1 to
4 m. A spectrum analyzer with peak detection capabilities is used to
fi
fi
find the maxima of the
radiated emissions during the testing. Then,
final measurements are taken using a spectrum
analyzer with quasi-peak function with a measurement bandwidth of 120 kHz. The test
setup is shown in Figure 4.4. The limits for radiated emissions per EN-55011 for group 1
devices are presented in Table 4.2.
In reality, devices to be tested are not usually taken directly to the open-
fi
fi
eld test site.
Rather, they are
ensive radiated emissions in a small shielded
room. The compliance testing is then conducted in the 10-m open-
fi
first scanned for potentially o
ff
eld test site, paying
special attention to peak emissions detected in the shielded room. This is almost a practi-
cal necessity, because open-
fi
eld test sites, even when located far from large metropolitan
areas, are still inundated by human-made RF signals. As an example, Table 4.3 shows
results obtained recently when testing an implantable-device programmer for radiated
emissions. The speci
fi
fi
c frequencies selected for testing were identi
fi
ed the night before
taking the device to an open-
eld test site in the middle of Texas's hill country. Figure 4.5
shows the device being tested at the open-
fi
eld site. The device sits atop a motorized
turntable. A biconical antenna can be seen placed 10 m away from the device under
test. At the 10-m distance speci
fi
fi
ed for the tests, radiated emissions have their electric-
fi
eld
1 TO 4 METERS
3 TO 30 METERS
EUT
COAXIAL CABLE
80 cm
NON-CONDUCTIVE TABLE
TURNTABLE
GROUND PLANE
eld test site. The device under test is placed on a noncon-
ductive table 0.8 m above a ground plane, and the distance between the device and the antenna is 10 m.
Figure 4.4
Setup for conducting radiated-emission measurements in an open-
fi