Biomedical Engineering Reference
In-Depth Information
Voltage-Measurement
Jaw
Heat-Shrink
Tubing
Nylon Bolt
and Nuts
Spring
Lead to J3
Dielectric
Separator
Lead to J2
Current-Injection
Jaw
Figure 3.22 An alligator clip can be converted into a Kelvin probe by replacing the standard metallic axis by a nylon bolt with nylon
spacers to isolate the jaws from each other. For the same reason, the ends of the inner spring must also be insulated.
suitable insulator. Two separate leads are then used to connect to the jaws of the probe, one
to inject current and the other to sense voltage.
Once the circuit and the probe are assembled, calibrate the adapter to produce exactly 1 A.
Plug the power cord of the instrument under test to the hospital-grade ac plug J1, and clip the
Kelvin probe to an exposed conductive point of the case that is supposed to be protectively
grounded. A digital voltmeter connected between J4 and J5 will directly read the protective
ground resistance on a scale of 1 V/
. It must be remembered, however, that the measure-
ment of resistance provided by this instrument only approximates the impedance test intended
by the standards. The discrepancy between the methods is especially evident for high-power
circuits, since a dc measurement of resistance does not convey any information regarding the
inductive component of impedance. Moreover, dc ohmmeters are usually fooled by the polar-
ized interface that results when an oxidation layer forms between connections in a defective
ground system. This last concern may be alleviated by running the test once again but with
the current injection polarity reversed. Nonlinear polarization indicating oxidation must be
suspected if resistance measurements taken with opposite current injection polarities do not
agree to a high degree. Failing this test is an immediate show-stopper. Before proceeding
with any further testing, you must locate the faulty connection responsible for compromis-
ing the integrity of the protective ground.
Measuring Leakage and Patient Auxiliary Currents
Leakage and auxiliary current tests are the most important tests to establish the electrical
safety of a medical electronic instrument. These are also the tests that are most commonly
failed during safety approval submissions as well as during the periodic tests that hospitals
conduct to ensure the safety of medical electronic devices throughout their service life. In
the case of medical electronic instruments, measurements of leakage and auxiliary currents
are taken using a load that simulates the impedance of a human patient. The AAMI load is
a simple RC network that presents an almost purely resistive impedance of 1 k
for fre-
quencies up to 1 kHz. As shown in Figure 3.23 a , this load constitutes the core of the cur-
rent measuring device. If a 1-
erent
frequencies, the high-impedance RMS voltmeter within the measuring device would read
the values presented in the graph of Figure 3.23 b .
A current is forced through the AAMI load at di
ff
Search WWH ::




Custom Search