Biomedical Engineering Reference
In-Depth Information
Fig. 8.9
Measured total VCDL delay versus control voltage V C
out, for the actual operation indeed performs image rejection via back-end digital
signal processing (Sect. 8.2.1 ). Using the measured input-referred noise and the
coil impedance (Sect. 8.2.2 ), we infer the receiver NF. In the palm system, a
passive gain of 28 offered by the resonance matching (Sect. 8.2.4 )lowersthe
NF from 22.5 to 0.9 dB. In the 1-chip system, a passive gain of 2.1 lowers the
NF from 6.1 to 2.2 dB. The combination of the small input-referred noise and
the resonance matching (optimum noise matching) leads to sufficiently low NF,
making the ping-pong-ball-sized magnet and the lossy on-chip coil viable system
options.
8.3.2
Transmitter Measurements
The measured output impedance of the differential PA (Sect. 8.2.3 ) is 27. With the
power supply V DD of 3.3 V, the maximum deliverable power at the fundamental
tone is 82 mW. The measured delay versus control voltage, V C , of the entire voltage-
controlled delay line (VCDL) in the duty cycle controller (Fig. 8.5 , bottom left) is
shown in Fig. 8.9 . The delay is altered from 29 to 2 ns, as V C is varied from 0 to
3.3 V.
The measured duty cycle as a function of V C for a 21-MHz excitation signal is
shown in Fig. 8.10 .AsV C is changed from 0 to 3.3 V, the duty cycle increases from
0 % to 45 %. This translates to the tuning of output power at the fundamental tone
from 0 to 80 mW (98 % of the total deliverable power of 82 mW).
Search WWH ::




Custom Search