Biology Reference
In-Depth Information
￿ Enter the plate calibration mode of the AxioVision Mark&Find module.
￿ Select the multiwell plate geometry template matching the microplate.
￿ Calibrate the plate geometry using the provided seven-point routine.
15.8.5 Step 5: adjust TIRF excitation
￿ Select the TIRF focus adjustment tool in the objective turret.
￿ Change the filter cube turret to the TIRF adjustment cube.
￿ Turn on the 642 nm laser and set the power to 5 mW.
￿ Observe the beamwith the AxioCamMRm and adjust the TIRF focus knob on the
TIRF3 slider to minimize the spot size.
￿ Select the TIRF angle adjustment tool in the objective turret.
￿ Observe the beam with the AxioCam and adjust the TIRF angle micrometer to
move the spot to a position approximately 20% close to the image bottom.
(The exact position has to be determined from a sample with fluorescent beads
and should be used to set this angle. Refer to the Zeiss TIRF3 manual.)
15.8.6 Step 6: sample for subpixel alignment of color channels
￿ Prepare a sample with immobilized 100 nm TetraSpec beads (Invitrogen) on
poly- D -lysine (PDL)-coated glass-bottom microplate dish (SensoPlate) or culture
dish (CellView) (Greiner).
￿
Image surface-immobilized beads with simultaneous laser (TIRF) excitation (488
and 561 nm). Use images to calculate image registration transformations
described in Step 11.
15.8.7 Step 7: alignment of the CRISP focus-stabilization
(autofocus) system
￿ The extension tube adjustment and prism adjustment typically have to be
performed at the installation of the CRISP unit only.
￿ Start the ASI Console software, connect to the CRISP controller, and start the
three-step adjustment routine.
￿ Manually optimize the detector lateral adjustment to maximize the dithering
difference signal on the ASI controller.
￿ Finish the adjustment routine with the Gain Setting step.
￿ Generate the Focus Curve using the control button and check if the vertical axis is
in the linear part of the curve and well separated from the extremes.
15.8.8 Step 8: preparation of antibleaching and antiblinking
imaging buffer system
15.8.8.1 Materials
Trolox, (
)-6-hydroxy-2,5,7,8-tetramethylchromane-2-carboxylic acid (Aldrich
238813, 97%, MW
¼
250.29 g/mol)
Search WWH ::




Custom Search