Biomedical Engineering Reference
In-Depth Information
Intended use of the device
Measurements and measuring instruments to be used
Sterility requirements
Lifetime of the device
In addition one should consider the following joint implant system criteria:
a) Medical and Engineering Criteria
Material and wear product compatibility
Adequate mechanical strength
Minimization of the joint reaction forces
Minimization of fixation interface tension
Avoidance of fixation interface shear
Uniformity of interface compression
Duplication of anatomical function
Adequate fit for the patient population
Manufacturability
Inventory costs
b) Medical and Surgical Criteria
Treatment of a broad variety of pathologies
Maximal preoperative options
Maximal intraoperative options
Maximal postoperative options in case of failure
Salvagability
Tolerance for misalignment
Ease of implantation
3.2.3.6 Verification
Verification of the design should be performed at designated phases of the design
process, to verify that the design output is consistent with the required input, using
methods appropriate to the design as determined by the Design Committee and
defined in the Design Plan. These methods may include:
Determination of motion
Stress analysis
Simulator testing for loading and wear
Manufacture and testing of prototypes
Testing of materials
Biocompatibility studies
Comparison of the proposed design to similar approved
devices on the market, including any clinical investigations
Search WWH ::




Custom Search