TABLE 1.4
Aerodynamic Cutoff Diameters ( μ m) for Instruments That Determine Particle Size by Inertial Impaction as Reported by the 
Manufacturer
Twin-Stage 
Liquid 
Impinger
Andersen 
Viable 
Impactor
Andersen Nonviable 
Impactor
Next Generation 
Impactor
Marple-Miller 
Impactor
Apparatus
USP B
Multistage Impinger
Delron
Flow Rate
(L/min)
Stage
60
60
30
60
80
100
28.3
28.3
60
90
30
60
100
30
60
90
12.5
Pre-separator
15.0
13.0
10.0
−1
8.6
8.0
0
6.4
9.8
7.1
9.0
6.5
6.5
1
16.9
13.3
11.8
10.4
4.7
5.8
4.4
5.2
11.0
7.8
6.0
10.0
10.0
8.0
11.2
2
9.3
6.7
5.6
4.9
3.3
4.7
3.3
3.5
6.6
4.6
3.6
5.0
5.0
4.0
5.5
3
4.5
3.2
2.7
2.4
2.1
3.3
2.0
2.6
3.9
2.7
2.1
2.5
2.5
2.0
3.3
4
2.5
1.7
1.4
1.2
1.1
2.1
1.1
1.7
2.3
1.6
1.2
1.25
1.25
1.0
2.0
5
0.65
1.1
0.54
1.0
1.4
0.96
0.72
0.63
0.63
0.6
0.9
6
0.7
0.25
0.43
0.84
0.57
0.42
0.5
7
0.4
0.51
0.33
0.23
MOC
0.31
0.13
0.055
Source: Dunbar, C. et al., KONA , 16, 7, 1998.