FIGuRE 13.10 Correction.of.aberrations.induced.by.250.μm.thick.ixed.mouse.brain.slices..(a-c).Lateral.( x - y ).
and.axial.( x - z .and. y - z ).images.of.a.single.1.μm.diameter.bead.under.a.brain.slice,.with.only.system.correction.(a),.
with.a.4.×.display.gain.(b).and.ater.full.AO.correction.(c)..(d).Lateral.intensity.proiles.along.the.light.gray.and.dark.
gray.lines.in.(a).and.(c)..(e).he.corrective.wavefront,.ater.subtraction.of.system.aberrations,.in.units.of..excitation.
light. wavelength. (850. nm),. using. 12. independent. subregions. and. direct. phase. measurement.. Scale. bars:. 2. μm..
(Ji, N.,.et al.,. Nat. Meth .,.7,.2,.141-147,.2010.)
biological. applications,. it. would. be. desirable. if. the. intrinsic. luorescence. from. the. sample,. rather.
than. a. guide. star. like. a. luorescent. bead,. can. be. used. for. AO. correction.. Instead. of. using. the. cen-
troid. shit. of. a. point. object,. such. as. the. luorescent. bead,. to. measure. the. beam. delection,. one. can.
use. image. correlation. to. measure. the. image. shit. when. diferent. pupil. segments. are. turned. on.. In.
Figure.13.11 , .luorescent.images.of.several.neurons.at.the.bottom.of.a.ixed.brain.slice.were.used.to.
measure.the.aberration.caused.by.a.cover.slip.and.the.300-μm-thick.brain.tissue..hree.sets.of.image.
data. are. shown—without. correcting. the. system. aberration,. with. system. aberration. correction,. and.
with.correction.for.both.the.system.and.sample-induced.aberration—and.the.image.quality.improves.
progressively.. (It. is. not. always. true. that. correcting. the. system. aberration. improves. image. quality..
Depending.on.the.nature.of.the.aberrations,.system.aberration.may,.in.some.cases,.partially.cancel.
the. sample-induced. aberration.. As. a. result,. correcting. the. system. aberration. may. actually. reduce.
the. image. quality.). he. improvement. is. especially. striking. in. the. axial. plane,. because. axial. resolu-
tion.is.usually.more.susceptible.to.aberration.than.lateral.resolution—a.long-standing.observation.in.
microscopy.[15].that.can.be.simply.explained.with.our.focus.model..For.most.of.the.light.rays.within.
the.excitation.beam,.as.the.rays.start.to.deviate.from.and.miss.the.focus,.the.volume.they.form.has.a.
longer.dimension.axially.than.laterally.
In.addition.to.ixed.in.vitro.samples,.AO.can.be.used.to.improve.imaging.quality.in.vivo..Using.the.
pupil-segmentation-based. AO,. we. systematically. characterized. and. corrected. the. optical. aberrations.
 
 