Biomedical Engineering Reference
In-Depth Information
FIGURE 12.3
Ultrasound calibration. A tracked A-mode probe (a) has an associated 3D coordinate system.
The calibration challenge is to relate the position along the A-mode time axis (b) to a point
in 3D. For the B-mode probe (c), we want to relate 2D coordinates ( u , v ) in the B-mode scan
(d) to 3D space.
where u is the speed of sound in tissue, P 0 is the position of the transducer,
and is the unit vector along the beam direction. The calibration process
involves finding P 0 and A method to achieve this using multiple scans of
a steel ball is described by Maurer et al. 40
B-mode calibration relates a 2D image pixel, ( u , v ), to a 3D point, P , with
respect to a tracker on the probe. Prager 41 describes the transformations as
follows:
ˆ
ˆ .
P
T C T T T R T R P x
(12.2)
T ,
where x
T R P is the rigid transformation from x to the
probe tracker, T T R is from probe tracker to tracking device coordi-
nates, and T C T transforms tracking coordinates to the reconstruction or
(
s u u , s v v ,0,1
)
Search WWH ::




Custom Search