Environmental Engineering Reference
In-Depth Information
14
13
Sensor
Sensor
1
3
5
7
9
11
13
15
2
4
6
8
10
12
14
16
12
11
10
Trend
9
8
10 6
1
10
100
1000
10,000
100,000
Suction, kPa
Figure 4.40 Measured temperature changes for 16 FTC-100 thermal conductivity suction sen-
sors (after Hu et al., 2007).
4.2.9.3 Calibration Equations for Thermal Conductivity
Suction Sensors
The electrical response from a thermal conductivity suction
sensor can be plotted versus the logarithm of soil suction.
The resulting curve is S shaped with anchor-type points at
minimum suction (i.e., assumed zero suction) and a maxi-
mum suction (i.e., near 10 6 kPa). The calibration curve has
a sigmoidal shape.
In 1999, Feng and Fredlund proposed the following form
for the calibration equation for thermal conductivity suction
sensors:
The a 1 and c 1 fitting parameters in Eq. 4.6 have the same
meaning as the a and c parameters in Eq. 4.5. The fitting
parameter b 1 is the suction when T is at the inflection
point of the calibration curve. In other words, it is midway
between the “dry” reading and the “saturated” sensor read-
ing. Stated another way, b 1 is equal to the soil suction at
the inflection point (i.e., suction when T = (a 1 + c 1 )/ 2).
Then, d 1 is the slope of the calibration curve at point b 1
(i.e., dψ/dT | T = ((a 1 + c 1 )/ 2 ) ) (see Fig. 4.41).
The Hu et al. (2007) revision to the calibration equation
was used to analyze the calibration data of 16 FTC-100
thermal conductivity suction sensors. The calibration mea-
surements are summarized in Table 4.5. The Shuai et al.
(2002) temperature correction was applied to the measured
data prior to determining the best-fit equation parameters.
The average values for the four calibration parameters on
the 16 sensors are as follows: a 1 =
b (T a)
c T
ψ =
(4.5)
where:
ψ
=
applied suction, kPa,
9 . 19 C, b 1 =
temperature rise, C, and
261 . 1 kPa,
T
=
185 . 0 kPa/ C. The a 1 parameter is
the thermal conductivity sensor reading when the ceramic
is saturated with water, and c 1 is the thermal conductivity
sensor reading when the ceramic is dry. The b 1 variable
is the soil suction corresponding to a temperature change
halfway between dry and water saturated conditions (i.e., the
inflection point along the calibration curve on a logarithm
scale). The d 1 variable is the slope corresponding to point b 1 .
The following points are noteworthy from the calibration
of the 16 thermal conductivity suction sensors. First, the a 1
and c 1 parameters provide the most consistent measurements
on the calibration curve. The standard deviation is similar
for both variables with values of approximately 0.3-0.4 C.
A typical value for the a 1 parameter was 9.2 C and a typi-
cal value for the c 1 parameter was 13.2 C. These values are
primarily a function of the constant current applied during
calibration. The b 1 parameter provides an indication of the
range where the sensor has the greatest sensitivity to suc-
tion changes. The 16 suction sensors calibrated show the b 1
parameter to be around 260 kPa. The b 1 and d 1 parameters
13 . 3 C, and d 1 =
c 1 =
a , b, c, d
=
four parameters to define the shape of the
calibration curve.
Of the four fitting parameters, a and c are equivalent to
the temperature rise T for the thermal conductivity sensors
under saturated and dry conditions, respectively. The parame-
ter b defines a particular horizontal position on the calibration
curve and the parameter d defines the slope of the curve
at point b . However, point b was not the most appropriate
point around which to perform the best-fit regression analysis.
Rather, it is better to use the inflection point on the semiloga-
rithm calibration curve for the definition of the b parameter.
In 2007, Hu et al. proposed a modification to Eq. 4.5 that
would give more significant physical meaning to each of the
fitting parameters. The new proposed calibration equation
was written as
T a 1
c 1 T
d 1 c 1 a 1
4 b 1
ψ = b 1
(4.6)
 
Search WWH ::




Custom Search