1000
Fig. 13
Comparison of the FE and mean measured principal strains for each gauge location and
for all load configurations with support in incisor teeth
FE = 0.9374EXP - 5.3918
600
R
2
= 0.9786
400
200
EXP (microstrain)
0
−
1000
−
500
500
−
200
−
400
−
600
−
800
Fig. 14
Linear regression analyses were performed to determine the overall correspondence
between FE and mean measured strains in all supports
) and the RMSE value of the measured strains is
1%. By separating the minimal strains from the maximal strain, relevant differences
can be observed. The other load configurations present similar results [
58
].
This study proved that FE models of the mandible can reproduce experimental
strains within an overall agreement level of 10%. The FE models correctly repro-
duced bone strains under different load configurations. For this reason, the FE model
The intercept value is small (6
"