Fig. 14
(
a
,
top
): Hydrophobicity (log
D
) vs. pH curves for penciclovir and vidarabine. (
b
,
middle
):
Less internal standard response variation was observed while using vidarabine as the internal stan-
dard (CV = 13.02 %) but 43 % of the calibration standards (CS) and quality controls (QC) were
rejected. Extraction: MCX (mixed-mode strong cation exchange)-based solid-phase extraction.
(
c
,
bottom
): More IS response variation was observed while using a deuterated internal standard,
penciclovir-d
4
(CV = 23.81 %) but 100 % of the CS and QC samples were accepted. Reproduced
from ref. [
36
] with permission from Elsevier