Table 2 MRM transitions, cone voltages (CV), and collision energies (CE) used in the analysis of
the antidepressants and the deuterated internal standards included in the LC-MS/MS method
Compound
CV
Transition
CE
Venlafaxine
25
278.1 > 57.8
278.1 > 260.3
18
12
Citalopram
35
325.1 > 109.1
325.1 > 262.1
26
20
Desipramine
25
267.0 > 43.8
267.0 > 71.8
40
15
Imipramine
25
281.1 > 57.6
281.1 > 85.9
40
16
Paroxetine
35
330.0 > 69.7
330.0 > 192.3
30
20
Nortriptyline
25
264.2 > 90.8
264.2 > 233.3
22
16
Fluvoxamine
25
319.0 > 70.8
319.0 > 86.8
18
18
Amitriptyline
30
278.1 > 90.8
278.1 > 233.2
24
18
Nor fl uoxetine
15
296.0 > 29.8
296.0 > 134.1
10
6
Sertraline
18
306.0 > 159.1
306.0 > 275.1
26
12
Fluoxetine
22
310.1 > 43.7
310.1 > 148.2
12
8
Norclomipramine
25
301.1 > 71.8
301.1 > 270.2
18
16
Clomipramine
32
20
Imipramine-3 25 284.1 > 89.0 15
Paroxetine-d6 35 336.1 > 76.0 30
Nortriptyline-d3 30 267.1 > 233.3 15
Nor fl uoxetine-d6 15 302.2 > 140.1 6
Fluoxetine-d6 25 316.1 > 43.7 12
Clomipramine-d3 30 318.1 > 88.9 18
Underlined transitions were used as quantifiers ( Source : From de Castro A et al. (2008) J Pharm
Biomed Anal. 48:183. With permission)
22
315.0 > 57.8
315.0 > 85.9
6.3.2
Calibration Model
Calibration curves ( n = 6, six different days) were generated at eight concentration
levels in the range from 2 to 500 ng/mL and 2-1,000 ng/mL for oral fluid and
plasma, respectively. Coefficient of determination ( r 2 ) was >0.99 for all analytes
using a 1/ x weighted linear regression, except for fluvoxamine, for which a
quadratic response was observed.
 