Fig. 5 Typical chromatograms showing injection volume ( V i nj ) optimization for plasma samples
spiked with 50 pg/mL of TACA and DEX-AC and extracted by selective SPE. The V inj was ( a ) 0.1 m L;
( b ) 0.5 m L; ( c ) 2 m L; ( d ) 5 mL. A 15 cm by 300 mm I.D. capillary column was used for separation; the
manufacturer-recommended V inj was 0.1-0.2 mL. Selective SPE and two-segment gradient separation
enabled a large V inj on a 300 mm I.D. capillary column without deterioration of chromatographic
performance and with greatly increased sensitivity (Reproduced with permission from American
Chemical Society)