Figure 7.9
Magnetic sensor trajectory projection showing that the motion
of the robotic camera was triggered inside the designed
detection range [16].
This small reaction range may be explained by the location of P 0
near the silicone model entrance, but the small reaction range in P 5 ,
0.21 mm, can be explained by one of the following reasons:
(i) A local distortion in the magnetic ield appeared between the
LUT construction and the robot manipulation.
(ii) A change on silicone vasculature lumen shape between the
LUT construction and the robot manipulation.
(iii) Excess of speed in the motion of the catheter.
Table 7.3
Comparison of deined and experimental detection ranges for
all reference points in M 0 and system setting time
Reference
point
Maximum
deined
detection
range
(mm)
Experimental
reaction
range (mm)
Setting
time (s)
Step count
DOF1
DOF2
DOF3
P 0
10
4.56
9.6
27718
14812
1456
P 1
10
7.00
3.2
2836
8816
2342
P 2
10
6.68
0.4
0
0
1104
P 3
10
8.80
2.4
6334
1950
0
P 4
10
7.81
1.4
3740
0
218
P 5
10
0.21
2.2
5652
3396
0
P 6
10
7.03
2.6
9156
650
0
 