Biomedical Engineering Reference
In-Depth Information
T represents the light transmission through the sample and is given by P t /P 0 , and SO 2 is
defined as c HB /(c HB þ
c HBO 2 ).
The measurement of SO 2 in blood can be performed either in vitro or in vivo. In vitro
measurement using a bench top oximeter requires a sample of blood, usually withdrawn
from a peripheral artery. The sample is transferred into an optical cuvette (a parallel-wall
glass container that holds the sample), where it is first hemolyzed and then illuminated
sequentially by light from an intense white source after proper wavelength selection using
narrow-band optical filters.
SO 2 can also be measured in vivo using a noninvasive pulse oximeter. Noninvasive opti-
cal sensors for measuring S a O 2 by a pulse oximeter consist of a pair of small and inexpen-
sive light-emitting diodes (LEDs)—typically a red (R) LED around 660 nm and an infrared
(IR) LED around 940-960 nm—and a single, highly sensitive silicon photodetector. These
components are typically mounted inside a reusable spring-loaded clip or a disposable
adhesive wrap (Figure 10.30). Electronic circuits inside the pulse oximeter generate digital
switching signals to turn on and off the two LEDs in a sequential manner and synchronously
FIGURE 10.30 (a) Transmission-type pulse oximeter finger probe, and (b) disposable finger sensor. Courtesy of
Nellcor Puritan Bennett, Inc., Pleasanton, CA.
Search WWH ::




Custom Search