Biomedical Engineering Reference
In-Depth Information
flow direction
outlet
reference electrode
cell culture medium
test substances
immobilized cells
pH
pO 2
IDES
electronic module
sealing element
reference
electrode
reference
electrode
amperometric
sensor
amperometric
sensor
earthing
electrode
earthing
electrode
pH electrode
pH electrode
FIGURE 10.6 Schematic drawings of a lab-on-a-chip system. (Top) Multisensor chip consists of pH,
pO2, and conductivity (impedance) electrodes incorporated in a microfl uidic cell. (Bottom) The layout of
electrodes on the chip. (Reproduced from [91], with permission from Elsevier.)
offering enhanced measurement capabilities for many clinical applications [123]. Such
smart sensors can even be used in closed-loop controlled medical devices [14, 124].
The commercially available handheld i-STAT system (Abbott Laboratories) is a typi-
cal example of such
TAS or lab-on-chip devices (www.i-stat.com). The self-contained
test cartridge consists of microfabricated thin fi lm multisensors with a pH sensor on
silicon platform, on-chip reference electrodes, washing and calibration solutions, and
fl uid transportation system. Together with their handheld electrochemical analyzer and
appropriate cartridges, the system is capable of performing a comprehensive panel of
critical tests for blood chemistries, coagulation, hematology, glucose, or cardiac mark-
ers. The chemical sensors and biosensors in the cartridge provide good diagnostic char-
acteristics such as high sensitivity and reproducibility, short analysis time, and a small
sample size requirement [125].
ยต
 
Search WWH ::




Custom Search