Biomedical Engineering Reference
In-Depth Information
Table 5.5-3 Environmental use classification ranking
Table 5.5-4 Preventive maintenance ranking
Point score
Risk category IV: Environmental use
classification (EC)
Point score
Risk category III: preventive
maintenance (MR)
Primary area of equipment use
PM frequency
5
Anesthetizing locations
5
Monthly
4
Critical care areas
4
Quarterly
3
Wet locations/labs/exam areas
3
Semi-annually
2
General patient care areas
2
Annually
1
Non-patient care areas
1
Not required
Medical equipment management
program deletion
of a preventive maintenance or repair, and the number
of years beyond the American Hospital Association's
recommended useful life.
Medical equipment and devices that pose little risk and score
less than 6 points can be deleted from the management
program as well as from the clinical equipment inventory.
Future versions of this application will also consider
''dynamic'' risk factors such as user errors, mean-time-
between failure (MTBF), device failure within 30 days
Peer performance reviews
The virtual instrument shown in Figure 5.5-18 has been
designed to easily acquire and compile performance
Figure 5.5-18 Performance reviews using VI.