Biomedical Engineering Reference
In-Depth Information
Fig. 7.38 Boxplot for the
fractional-order values n
identified from the 31
volunteers
Fig. 7.39 The original Z PA R
impedance ( bold line )as
identified for patient “22”
from Table 7.15 and the
modified impedance ( stars )
with twofold increased R ts
(denoted in figure as “test
case”)
order parameter with frequency, we re-identify the ladder network model in the
range 5-50 Hz, and we obtained a new fractional-order value, depicted in Fig. 7.38
in the right-column. There was no statistically significant difference between the
two identified sets ( p< 0 . 7). This validates the expected theoretical result that the
fractional-order value is independent on the frequency.
Further on, the sensitivity of the fractional-order value to changes in mechanical
parameters has been evaluated by using Z PA R with modified R ts increased twofold,
and fitting Z REC to it. Figure 7.39 shows the original Z PA R impedance identified
for patient “22” from Table 7.15 and the modified impedance (denoted in figure as
“test case”). To the modified impedance (i.e. test-case impedance) we have fitted
the model of Z REC asshowninFig. 7.40 and obtained the following parameter val-
ues: R ua =
1 . 3 e 7
0 . 0017 (kPa s 2 /L); C ua =
(kPa s/L); L ua =
0 . 0034 (L/kPa); R
=
0 . 0005 (kPa s 2 /L); C =
0 . 098 (kPa s/L); L =
0 . 0016 (L/kPa); λ =
1 . 10; χ =
0 . 52;
α
0 . 03. The simulated increase in tissue resistance is re-
flected by an overall increase in the airway resistance value and modified recurrent
ratio values. The resulting fractional-order value is then calculated from the identi-
fied recurrent ratios using ( 5.38 ), it following that n
=
1 . 10 and total error E t =
=
0 . 19. This implies that our
proposed model is able to capture variations in the respiratory parameters and the
fractional-order value is a suitable parameter for classification.
Search WWH ::




Custom Search