Biomedical Engineering Reference
In-Depth Information
X
0
1
0
E
1
E
2
E
3
E
4
Figure 3.31
Illustration of a “don't-care” in electrode activation.
E
1
E
2
E
3
E
8
E
4
E
7
E
6
E
5
(a)
Electrode
12345678
Activation
Sequence
0
1
0
0
X
X
X
X
1
0
0
X
X
X
X
0
0
0
X
X
X
X
0
1
0
X
X
X
X
0
1
0
X
X
X
X
0
1
0
0
X
X
X
0
1
0
0
X
X
X
0
1
0
0
X
X
X
0
1
0
0
X
X
X
(b)
Figure 3.32
Example of activation sequence calculation: (a) routing and layout information; (b) calculated
activation sequences.
We use the three values “1,” “0,” and “x” to represent the electrode-activation
sequences for a bioassay. An example is shown in Figure 3.32. A droplet is
routed anticlockwise, one electrode per step, along the loop consisting of
eight electrodes. Suppose that at time instant (clock cycle)
t
0
, the droplet rests
on electrode E
2
. The activation sequence for each electrode is now calculated
and listed in Figure 3.32b.
In Figure 3.32b, each sequence contains several don't-care terms, which
can be replaced by “1” or “0.” By carefully replacing these don't-care terms,
the two activation sequences corresponding to E
1
and E
4
can be made identi-
cal. For example, we can map the four don't-cares in the activation sequence
for E
1
with “0010” and the four don't-cares in the activation sequence for E
4
with “0100.” We refer to such sequences as
compatible sequences
. Compatible