Digital Signal Processing Reference
In-Depth Information
VNA
DUT
R
Error Port1
Error Port2
DUT
R
Desired measurement
reference planes
Actual measurements are referenced internal to the VNA
and include cable, probe and connector losses and phase delays
Figure 9-35 Calibration removes the unwanted parts of the measurement, such as the
cable and probe effects.
a 1
b 2
b 1
a 2
Z L
Error Port1
Γ in
L
Figure 9-36 The simplest way to calibrate a VNA is to characterize the port error by
driving different loads ( Z L ): an open, a short, and a resistive load.
precision resistive load. For example, consider Figure 9-36, which represents
one port of a VNA driving a load Z L . The equations for the S -parameters can be
written in terms of the reflection coefficient at the load where the incoming wave
at port 2 ( a 2 ) is simply the reflected portion of the exiting wave ( b 2 ) a 2
= b 2 L .
b 1
= S 11 a 1
+ S 12 a 2
= S 11 a 1
+ S 12 b 2 L
b 2
= S 21 a 1
+ S 22 a 2
= S 21 a 1
+ S 22 b 2 L
S 21 a 1
b 2 ( 1
S 22 L ) = S 21 a 1
b 2
=
(9-51a)
1
S 22 L
S 21 a 1
b 1
= S 11 a 1
+ S 12
S 22 L L
(9-51b)
1