Digital Signal Processing Reference
In-Depth Information
herein taken to be approximately 1 Hz, allowing slight variations, and also
assuming that the peaks were Lorentzian. The line widths are proportional to
Im(ν
k
). It is emphasized that the mentioned smallest chemical shift difference
of≈2×10
−4
ppm is 4 times less than the typical line width of 8×10
−4
ppm.
The phases φ
k
(1≤k≤9) from complexvalued d
k
were all set to zero, so
that every d
k
becomes real, d
k
=|d
k
|.
Table 10.1
displays the input data for the normal breast tissue (upper panel
(i)), fibroadenoma (middle panel (ii)) and for breast cancer (bottom panel
(iii)).
The diagonal FPT
(−)
was used to analyze the FIDs with the coe
cients{p
−
r
,q
−
s
}of the polynomials P
−
K
and Q
−
K
computed by solving a system
of linear equations from chapter 4 (section 4.10) by treating the product in
G
N
(z
−1
)∗Q
−
K
(z
−1
) = P
−
K
(z
−1
) as a convolution. We solved the character
istic equation Q
−
K
(z
−1
) = 0 to extract the peak parameters.
As discussed,
this leads to K unique roots z
−
k
(1≤k≤K), so that the sought ω
−
k
is
deduced via ω
−
k
= (i/τ) ln (z
−
k
). The k th metabolite concentration C
−
k
was
computed from the absolute value|d
−
k
|of the reconstructed amplitude d
−
k
as
C
−
k
= 25|d
−
k
|C
ref
M/g ww.
We systematically increased the signal length for the same bandwidth (i.e.,
3 acquisition times) to demonstrate the constancy of the spectral parameters
for all three signals. The spectral parameters were examined at total orders
K = 500, 750 and 1000, where 2K = N
P
and N
P
denotes partial signal
length. Convergence occurred at K = 750 for all three FIDs, and remained
stable thereafter. We determined whether a given reconstructed resonance
was true or spurious by using the concept of Froissart doublets throughout
the whole Nyquist interval with a particular focus on the frequency range of
interest from 1.3 ppm - 3.3 ppm. For all three examined FIDs, of the 750
total resonances, 741 were identified as spurious Froissart doublets by their
zero amplitudes and the polezero coincidences; the nine remaining resonances
were genuine.
10.4.1 Pade-reconstruction of MRS data for normal breast
tissue
tissue. Three illustrative partial signal lengths (N
P
= 1000, 1500 and 2000)
are shown therein. At N
P
= 1000 (upper panel (i)), only eight of the nine
resonances were identified. In the interval where there should be two peaks,
PCho (k = 4) at 3.220012 ppm and PE (k = 5) at 3.220215 ppm, only a single
resonance was identified at 3.220189 ppm. Since this peak was closer to PE,
we gave it that assignment.
At N
P
= 1000, the reconstructed amplitudes and, therefore, the concen
trations for that single peak were approximately the sum of (PCho + PE).
At N
P
= 1000 all the reconstructed spectral parameters were fully exact for
lactate (Lac, k = 1) at 1.330413 ppm, alanine (Ala, k = 2) at 1.470313 ppm,
Search WWH ::
Custom Search