Chemistry Reference
In-Depth Information
Table 8.4 GC-MS Instrument Parameters for the Analysis of Fire Debris and
Accelerant Standards
GC Method Parameters
Mass Spectrometry Parameters
Temperature program: 50°C held for 1 min; increase to
225°C at 10°C/min and hold for 1.5 min
Split: 20:1
Total run time: 20 min
Injection port temperature: 250°C
Carrier gas: Heat a flow rate of 1.0 mL/min
MS transfer line temperature: 280°C
Ion source temperature: 250°C
Mode: positive ion
Full scan range: 50-250 Da
can be seen in the following examples.) From the sample of fire debris, the
chromatogram shown in Figure 8.30 was found.
Using the software available on the GC-MS system, it is possible to
extract certain ions from the chromatogram (see also Section 8.3.6). If, for
example, we are looking for the aromatic component in the chromatogram,
it would be appropriate to extract the ions 91, 105 and 119 Da. These ions
were extracted from the fire debris collected from the scene and are shown
in Figure 8.31.
Other classes of compounds also have associated ions that will be
extracted (examples are shown in Table  8.5). However, for the purposes of
this example, we shall use only the aromatic components. Initially, the chro-
matogram from the fire debris compared to the standard liquids will be con-
sidered (Figure 8.32).
550000
500000
450000
400000
350000
300000
250000
200000
150000
100000
50000
0
Time -->
1.00
2.00
3.00
4.00
5.00
6.00
7.00
8.00
9.00
10.0011.00
12.0013.00
14.00
15.0016.00
Time (minutes)
Figure 8.30 Total ion chromatogram from SPME-GC-MS analysis of fire debris.
Search WWH ::




Custom Search