Figure 1.1 (
continued
)
(c) Quadrupole ion trap constrains ions into the mass
analyzer that are then ejected to a detector based on their
m
/
z
values.
Essentially a single mass analyzer, ion trap can also function as a tandem
instrument by successively concentrating, fragmenting, and analyzing ions
(a process called tandem-in-time) in the physical mass analyzer. (d) In an ion
cyclotron resonance cell, trapped ions rotate with a specific frequency that is
proportional to their mass-to-charge ratio. (Reprinted, with permission, from
the
Annual Review of Biophysics and Biomolecular Structure,
vol. 33, © 2004
Annual Reviews,
www.annualreviews.org.)
opposite polarity. Quadrupole mass filter operates in a mass-selective
stability mode: by varying the amplitudes of RF and DC potentials,
quadrupole selects only the
m
/
z
value that will have a stable trajectory
through the filter and arrive at the detector, while all the other
m
/
z
values will have unstable trajectories and be removed from the filter.
Usually, quadrupole mass analyzer has a mass accuracy of 300 ppm.