Figure 1.12: The functional dependency between the axial resolution and the
ultrasound frequency for a rank of typical ultrasound speeds (see Table 1.1) in
biological tissue. The typical IVUS (30 MHz) frequency as well as the tolerance
in the axial resolution d r are emphasized.
1.4.1.2 Angular Resolution
Angular resolution is the capacity to discern two objects or events located in
the tangential direction [13, 14, 16] and depends on the beam width. The beam
Table 1.1:
Sound speed in selected
tissues [16]
Material
Sound speed (m/sec)
Fat
1460
Aqueous humor
1510
Liver
1555
Blood
1560
Kidney
1565
Muscle
1600
Lens of eye
1620
Average
1553
 