Biology Reference
In-Depth Information
Ions in the ICR cell irradiated with a pulse of RF at the same
frequency as ω c oscillate in stable ion cyclotron motions (Fig. 2c ).
Ions absorbing the energy increase the radius of their orbits induc-
ing thereby an oscillating charge in the walls of the ICR cell as they
precess. The frequency of this induced charge oscillation can be
detected and amplified. Because a wide variety of energies are
transmitted to the ICR cell, FT-ICR MS allows many ions to be
detected simultaneously. The advantages of detecting all ions
simultaneously derives from the “Fellgett advantage” [ 25 ] and is
derives from the expression that relates the orbital radius ( r ) with
the excitation electric field ( E 0 ) and the excitation time ( t excite ),
r = ( E 0 t excite )/2 B . The fact that the orbital radius of the excited ions
is independent of the m / z ratio, means that ions of different m / z
ratios can be excited to the same ICR radius. In addition, because
frequency is an easily and accurately measurable parameter, FT-ICR
MS has the highest potential for mass accuracy determinations.
Pivotal to any mass spectrometry experiment is the mass resolv-
ing power of the mass analyzer, m m , where Δ m is the full peak
width at 50 % of maximum peak height. Mass resolving power is a
measure of how well two closely spaced peaks can be resolved. In
FT-ICR MS | m m | = |ω/Δω|. Because the frequency of ion cyclo-
tron precession is directly proportional to the strength of the mag-
netic field, increasing the strength of the magnet increases
resolution. In addition, Fourier transform MS is unique in that
increased measurement time also increases both sensitivity and res-
olution [ 26 ]. Increased sensitivity is derived from the fact that ions
are not consumed during the detection process. In a zero-pressure
limit, the frequency-domain mass spectral peak width is given by:
(
)
212060
π
.
T
∆ω
=
where T is the time-domain data acquisition period [ 27 ]. It follows that
the mass resolving power is directly proportional to the observation
period. In addition, the resolution at low pressure is given by:
(
)
1 274 10 7
.
×
zB T
m
m
0
=
m
[ 28 , 29 ] clearly showing that the resolving power, m m , increases
linearly with the strength of the magnetic field ( B 0 ) and the acqui-
sition time. Thus, resolution can be improved either by increasing
the strength of the magnet (in teslas) or by increasing the detec-
tion duration. Mass resolution ( R ) exceeding 1,500,000 have been
achieved by FT-ICR MS. However, high-performance instruments
use superconducting magnets that are large in size and expensive
to maintain from the standpoint of cryogenic cooling.
Search WWH ::




Custom Search