Fig. 3.15 Interference by chloride in the autoanalyser determination of
phosphate
Source: Reproduced with permission from Elsevier Science [172]
Table 3.14 Dependence of chloride content of wash liquid on chloride content of sample
Chloride content of wash (mg L −1 )
Chloride content of sample (mg L −1 )
0
0-1800
4850
1800-7200
10000
7200-12600
15200
12600-18000
Source: Reproduced with permission from Elsevier Science [172]
related to the chloride concentration of the sample. The chloride content of the wash
solution need not be exactly equal to that of the sample. For chloride contents in the
sample up to 18 000mg L −1 , (ie seawater) the chloride concentration in the wash should
be within ±15% of that in the sample (Table 3.14). The use of saline standards is optional
but the use of saline control solutions is mandatory. Using good equipment, down to
0.02mg L −1 phosphate can be determined by such procedures. For chloride contents
above 18,000mg L −1 the chloride content of the wash should be within ±5% of that in the