to further fragment the product ions monitored
in Q3. In MRM 3 , the ion chromatogram is recon-
structed from fragments of the product ion in
the linear ion trap instead of from the product
ionasintheMRMmode,asshownin Figure 3 .
The MRM 3 mode further reduces background
interferences, which lead to lower LOQs.
It was demonstrated that MRM 3
LOD and LOQ using
five model proteins with
only a slight reduction of accuracy and preci-
sion compared to a conventional MRM assay.
The MRM 3
demonstrated its potential
for
biomarker veri
cation when it was applied to
quantitate PSA in plasma samples obtained
from three cancer patients using only a simple
solid phase extraction sample preparation step
targeted at PSA peptides. The low ng/ml
generally
resulted in three- to
fivefold improvements in
FIGURE 3 Schematic representation of MRM 3 on a hybrid triple quadrupole linear ion trap mass spectrometer with
extracted ion chromatogram peak for quantitation. (Reproduced with permission from reference #23).